Page 29 - 5010
P. 29
flow rate is conducted by means of valves 14 and 17, and regulation
of the pressure behind the core is conducted by means of valve 15).
measuring
1 – cylinder with air or hydraulic press; 2 – head capacity with oil; 3 – discharge capacity with
water; 4 – capacity with water (water pipe line); 5 – manifold; 6 – core holder with rock ample;
for
manometer
–
8
core;
the
of
front pressure behind the core; 9 – measuring capacity of the volume 10 cm 3 ; 10 – microreducer Figure 4.1 – Apparatus (installation) WCPE-1 (ВКТВ-1) for determining the
in
permeability coefficient of rocks (cores) for liquid (water).
pressure
measuring (control valve); 11 – 18 – valves.
for
manometer
–
7
4.4.5 After stabilizing water filtration through the core note the
readings on manometers (indicating pressure gages) in front of and
behind the rock sample and measure the volumetric water flow rate
28