instruments.  In  the  case  of  equipment  such  as  heavy  machinery
                           (traction,  compression,  hardness,  etc.),  scales,  air  conditioning
                           chambers,  etc.,  the  verification  can  only  be  done  on-site.  It  is  not
                           necessary  for  the  provider  to  have  their  own  laboratory  since  the
                           whole intervention is carried out onsite. However, the provider must
                           use working standards which are related to the calibration chains.

                               Calibration requirements
                               Several  problems  come  to  mind  when  thinking  of  calibration.
                           First of all, how can a particular measuring instrument be calibrated?
                           If it is a calliper, you will think about using gauge blocks. Has anyone
                           even considered measuring rods for a micrometer? What is to be done
                           with dynamometric spanners, balances, etc.? If you go into physical
                           chemistry,  etc.  it  gets  even  more  complex!  Some  methods  of
                           measurement demand equivalent methods of calibration. Fortunately,
                           some manufacturers of materials provide tips. When you look deeper
                           into the matter, you realize that quite often you talk about calibration,
                           but  what  you  actually  need  is  a  verification,  perhaps  even  a
                           metrological  confirmation  (see  ISO  10012  standard).  Therefore,  it
                           might  be  necessary  to  proceed to  an  internal  checking  between  two
                           interventions, which is just a simplified examination of good working
                           order. Calibration must be done intelligently, which means doing just
                           what  is  necessary;  it  is  not only  a  means  to  avoid  auditor’s  critical
                           views. How many firms, which work in mechanical engineering and
                           have their sets of gauge blocks calibrated in an accredited calibration
                           laboratory simply open their calibration certificate? Never must it be
                           forgotten  that  the  major  purpose  of  calibration  is  to  verify  the
                           measuring instrument and calculate the uncertainties that go with the
                           results of the measurements taken with that instrument. The question
                           of the interval of the calibrations inevitably arises quickly. The answer,
                           which  should  make  everybody  happy,  is  that  it  depends.  Some
                           methods of measurement meet a few demands, particularly in the field
                           of physical chemistry. In any event, measuring instruments should be
                           calibrated  reasonably  frequently,  so  as  to  detect  and  prevent  any
                           possible drift, but not too often because of the overall cost involved.
                               The reader should wary of any person who claims that they can
                           tell which intervals are the right ones. As a matter of fact, you always
                                                           77